ACC's Pyros Kinetix® tube reader offers a sensitivity as high as 0.001 EU/mL which can be achieved with Pyrotell®-T lysate, the most sensitive assay in the industry.
Product Description
The Pyros Kinetix® incubating tube reader is used with Pyrotell®-T lysate for performing kinetic turbidimetric endotoxin assays. Pyros Kinetix is designed for use with the Pyros® EQS software (software sold separately).
The Pyros Kinetix tube reader offers a 96 tube capacity. Each well is independently timed, allowing the operator to add more samples while a run is in progress. A sensitivity as high as 0.001 EU/mL can be achieved with Pyrotell-T lysate, the most sensitive assay in the industry. Note: Borosilicate reaction test tubes must be used in the Pyros Kinetix tube reader.
Product Applications
Pyros Kinetix® is designed for use with the Pyros® EQS software.
Product Benefits
- Maximum Sensitivity: As high as 0.001 EU/mL with Pyrotell®-T lysate
- Maximum Efficiency: The instrument has a 96 tube format and allows sample addition, even after the run has been initiated
- Ease of Use: The instrument uses depyrogenated 8 x 75 mm glass test tubes which minimize the occurrence of contaminated wells, an issue frequently associated with plastic microplates
- Flexible Testing: Variable volumes and ratios can be utilized
- Reduced Lysate Usage: As little as 50 μL per test
Product Features
- Precise Temperature Control—incubator temperature is held to 37°C ± 0.5°C
- Solid State Design—no moving parts, low maintenance
- Low Cost Accessories
Product System Specifications
Capacity - 96 reaction test tubes
Power Requirement - 100–250 V, 50/60 Hz
Light Source - LED
Dimensions - 18 w x 3.75 h x 16 d in. (45.7 x 9.5 x 40.6 cm)
Weight - 27 lbs., 5 oz. (12.5 kg) (without external power supply)
Temperature Range - 37°C ± 0.5°C
Computer Requirements - Intel® Pentium® 4 PC; 2.4 GHz Windows® XP SP2 OS; 1 GB RAM (if stand alone); 40 GB HD, 800 x 600, 256 Color Monitor; serial port or compatible USB adapter required (for reader communication)
Data Interface - RS-232C
Data Transmission Format - ASCII character strings, 9600 baud
Registration -
Instrument: CE EN 61010-1, EN 61326-2 1998
Power Supply: IEC601-1, IEC950, UL2601-1,UL1950, UL544, EN60950,
EN60601-1, CSA22.2#125, CSA22.2#601-1, CSA22.2#950. TUV D303673901
(Cert# B970924072006)
Pyros Kinetix® |
|||
#PKX02 |
Pyros Kinetix® Incubating Kinetic Tube Reader | ||
|
|||
风险提示:丁香通仅作为第三方平台,为商家信息发布提供平台空间。用户咨询产品时请注意保护个人信息及财产安全,合理判断,谨慎选购商品,商家和用户对交易行为负责。对于医疗器械类产品,请先查证核实企业经营资质和医疗器械产品注册证情况。
- 作者
- 内容
- 询问日期
文献和实验涡旋混匀基于偏心原理,用于混合试管中的液体组分、重悬细胞。 涡旋混匀仪的操作使用步骤: 将放置在平稳的工作台上,并确保其电源线接地良好; 打开电源开关; 根据需要,选择合适的转速; 按下启动按钮,开始工作; 在混匀过程中,可以根据需要适时监控混匀的程度,并根据需要调整混匀的转速; 混匀结束后,按下停止按钮,关闭电源开关。 涡旋混匀仪的维护保养方法: 使用完毕后,将电源开关关闭,并拔掉电源插头; 使用干净的湿布或纸巾擦拭表面,清除残留液体和杂质; 定期清理工作室,清除堆积的杂质
1、真空控制器工作原理是啥?不借助压力传感器怎么实现压力控制的? 当系统内真空压力高于或低于设定压力时,控制器内的压力感应器立即动作,使控制器内的开关触点断开或接通,此时设备停止工作;当系统内的真空压力回到设备设定的压力范围时,控制器的压力立即复位,使控制器内的开关触点接通或断开,此时设备正常工作;不借助压力传感器,无法实现压力控制。 2、竖直型冷凝器和倾斜型冷凝器, 哪种比较好?对于应用要求来说两者的冷凝效果相当。这两种冷凝器的冷凝面积和冷却效率是一样的。 3、旋转蒸发仪和真空
旋蒸的基本原理就是减压蒸馏,在减压的条件下,连续蒸馏大量易挥发性溶剂,以达到分离和纯化的目的。 旋蒸是研发及分析实验中用于浓缩干燥回收产物的一款必备的基本仪器。即:在减压条件下,将旋转蒸发瓶置于水浴锅中,边旋转边加热,增大蒸发面积(溶剂在烧瓶内壁形成一层液体薄膜)利于瓶内溶液蒸发,样品旋转产生的作用力可以有效的抑制样品的沸腾。 影响旋转蒸发仪蒸发效率的四要素:真空度、转速、水浴温度、冷凝器 1、真空度 影响系统真空最关键且会变化的因素是:真空泵、密封圈和真空管。 密封圈:作为承接蒸发管和冷凝
技术资料暂无技术资料 索取技术资料