    大鼠免疫循环复合物(rat CIq) ELISA试剂盒



    S
    ample
    collection and storages

    Serum - Use a serum separator tube and allow samples to clot for 30 minutes before centrifugation for 10 minutes at approximately 3000×g. Remove serum and assay immediately or aliquot and store samples at -20℃ or -80℃.Avoid repeated freeze-thaw cycles
    Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 30 minutes at 3000×g at 2-8℃ within 30 minutes of collection. Store samples at -20℃or -80℃. Avoid repeated freeze-thaw cycles.
    Cell culture supernates and other biological fluids - Remove particulates by centrifugation and assay immediately or aliquot and store samples at -20℃or -80℃. Avoid repeated freeze-thaw cycles.
    Note: The samples should be centrifugated adequately and no hemolysis or granule was allowed.
    Materials required but not supplied
    1. Standard microplate reader(450nm)
    2. Precision pipettes and Disposable pipette tips.
    3. 37 ℃ incubator
    Precautions
    1. Do not substitute reagents from one kit to another. Standard, conjugate and microplates are matched for optimal performance. Use only the reagents supplied by manufacturer.
    2. Do not remove microplate from the storage bag until needed. Unused strips should be stored at 2-8°C in their pouch with the desiccant provided.
    3. Mix all reagents before using.
    Remove all kit reagents from refrigerator and allow them to reach room temperature ( 20-25°C)
    Materials supplied
    Name 96 determinations 48 determinations
    Microelisa stripplate 12*8strips 12*4strips
    Standard 0.3ml 0.3ml
    Sample diluent 6.0ml 3.0ml
    HRP-Conjugate reagent 10.0ml 5.0ml
    20X Wash solution 25ml 15ml
    Chromogen Solution A 6.0ml 3.0ml
    Chromogen Solution B 6.0ml 3.0ml
    Stop Solution 6.0ml 3.0ml
    Closure plate membrane 2 2
    User manual 1 1
    Sealed bags 1 1
    Note: Standard concentration was followed by:
    201052.51.250 ng/mL.
    Reagent preparation
    20×wash solution:Dilute with Distilled or deionized water 1:20.
    Assay procedure
    1. Prepare all reagents before starting assay procedure. It is recommended that all Standards and Samples be added in duplicate to the Microelisa Stripplate.
    2. Add standard: Set Standard wells, testing sample wells. Add standard 50μl to standard well.
    3. Add Sample: Add testing sample 10μl Then add sample diluent 40μl to testing sample well; Blank well doesnt add anyting.
    4. Add 10l of HRP-conjugate reagent to each well, cover with an adhesive strip and incubate for 60 minutes at 37°C.
    5. Aspirate each well and wash, repeating the process four times for a total of five washes. Wash by filling each well with Wash Solution (400μl) using a squirt bottle, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Solution by aspirating or decanting. Invert the plate and blot it against clean paper towels.
    6. Add chromogen solution A 50μl and chromogen solution B 50μl to each well. Gently mix and incubate for 15 minutes at 37°C. Protect from light.
    7. Add 50μl Stop Solution to each well. The color in the wells should change from blue to yellow. If the color in the wells is green or the color change does not


    appear uniform, gently tap the plate to ensure thorough mixing.
    8. Read the Optical Density (O.D.) at 450 nm using a microtiter plate reader within 15 minutes.
    Calculation of results
    1. This standard curve is used to determine the amount in an unknown sample. The standard curve is generated by plotting the average O.D. (450 nm) obtained for each of the six standard concentrations on the vertical (Y) axis versus the corresponding concentration on the horizontal (X) axis.
    2. First, calculate the mean O.D. value for each standard and sample. All O.D. values, are subtracted by the mean value of the zero standard before result interpretation. Construct the standard curve using graph paper or statistical software.
    3. To determine the amount in each sample, first locate the O.D. value on the Y-axis and extend a horizontal line to the standard curve. At the point of intersection, draw a vertical line to the X-axis and read the corresponding concentration.
    4. Any variation in operator, pipetting and washing technique, incubation time or temperature, and kit age can cause variation in result. Each user should obtain their own standard curve.
    5. The sensitivity by this assay is 0.1 ng/mL.
    6. Standard curve


    Storage2-8.
    validity six months.

    FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE BEGINNING!



    试剂盒组成
    名称 96孔配置 48孔配置 备注
    微孔酶标板 12孔×8条 12孔×4条
    标准品 0.3mL 0.3mL
    样本稀释液 6mL 3mL
    检测抗体-HRP 10mL 5mL
    20×洗涤缓冲液 25mL 15mL 按说明书进行稀释
    底物A 6mL 3mL
    底物B 6mL 3mL
    终止液 6mL 3mL
    封板膜 2张 2张
    说明书 1份 1份
    自封袋 1个 1个
    注:标准品浓度依次为:84、2、1、0.5、0 ng/ml.
    试剂的准备
    20×洗涤缓冲液的稀释:蒸馏水按1:20稀释,即1份的20×洗涤缓冲液加19份的蒸馏水。
    洗板方法
    1. 手工洗板:甩尽孔内液体,每孔加满洗涤液,静置1min后甩尽孔内液体,在吸水纸上拍干,如此洗板5次。
    2. 自动洗板机:每孔注入洗液350μL,浸泡1min,洗板5次。
    操作步骤
    1. 从室温平衡60min后的铝箔袋中取出所需板条,剩余板条用自封袋密封放回4℃。
    2. 设置标准品孔和样本孔,标准品孔各加不同浓度的标准品50μL;
    3. 待测样本孔先加待测样本10μL,再加样本稀释液40μL;
    4. 随后标准品孔和样本孔中每孔加入辣根过氧化物酶(HRP)标记的检测抗体100μL,用封板膜封住反应孔,37℃水浴锅或恒温箱温育60min。
    5. 弃去液体,吸水纸上拍干,每孔加满洗涤液,静置1min,甩去洗涤液,吸水纸上拍干,如此重复洗板5次(也可用洗板机洗板)。
    6. 每孔加入底物A、B各50μL,37℃避光孵育15min。
    7. 每孔加入终止液50μL,15min内,在450nm波长处测定各孔的OD值。
    结果判断
    绘制标准曲线:在Excel工作表中,以标准品浓度作横坐标,对应OD值作纵坐标,绘制出标准品线性回归曲线,按曲线方程计算各样本浓度值。

    试剂盒性能
    1. 准确性:标准品线性回归与预期浓度相关系数R值,大于等于0.9900。
    2. 灵敏度:最低检测浓度小于0.1 ng/ml
    3. 特异性:不与其它可溶性结构类似物交叉反应。
    4. 重复性:板内变异系数小于10%、板间变异系数小于15%。
    5. 贮藏:2-8℃,避光防潮保存。
    6. 有效期:6个月
    免责声明
    1. 试剂盒仅供研究使用,不得用于临床实验或人体实验,否则所产生的一切后果,由实验者承担,本公司概不负责。
    2. 严格按照说明书操作,实验者违反说明书操作,后果由实验者承担。

    风险提示:丁香通仅作为第三方平台,为商家信息发布提供平台空间。用户咨询产品时请注意保护个人信息及财产安全,合理判断,谨慎选购商品,商家和用户对交易行为负责。对于医疗器械类产品,请先查证核实企业经营资质和医疗器械产品注册证情况。

    图标文献和实验
    相关实验
    • 循环免疫复合物(CIC)的检测方法

      循环免疫复合物(CIC)的检测方法大致可分为两类,即抗原特异性方法和抗原非特异性方法。前者通过区别游离的抗原和与抗体结合的抗原,选择性测定含有某种特定抗原的IC。在已知由某种抗原引起的免疫病理反应的疾病中,可应用此类方法。抗原非特异方法,则不考虑形成IC抗原的性质,根据免疫球蛋白分子在结合抗原以后发生的物理学和生物学特性的改变进行检测。由于体内形成的CIC可能涉及多种抗原—抗体系统。所以临床上多采用此法。其检测方法种类繁多。大致可归纳为:1、物理方法 根据IC分子量大小,表面电荷和溶解度等特性

    • 特异性循环免疫复合物的测定

      特异性CIC分为单特异性CIC和双特异性CIC两类。前者适用于检测已知抗原及其相应抗体组成的CIC。双特异性CIC是指对组成CIC的抗原、抗体和补体中某两种成分组合明确的CIC,常采用的检测方法为ELISA和RIA技术,需要两种特异性各异的抗体。因此,检测要求条件较高,但能较全面地反映CIC的病理意义。 在检测双特异性CIC时,由于要求有两种特异性抗体,每种抗体均可用作包被或检测抗体,故具体到检测某类双特异性CIC时,根据两种抗体的不同,可使用两种方法。 以ELISA法为例,如检测

    • 脑梗死与循环免疫复合物的相关性研究

      摘要 目的 探讨循环免疫复合物与脑梗死的关系。方法 采用酶联免疫吸附试验(ELISA)方法测定脑梗死患者血中巨细胞病毒抗体(CMV-IgG与CMV-IgM)及衣原体抗体(CP-IgG),并与健康对照组比较。结果 脑梗死组循环免疫复合物显著升高(P﹤0.01),其中CMV-IgM在发病早期明显升高(P﹤0.01),随着病情康复,其水平逐渐下降。CMV-IgG和CP-IgG也呈现类似变化。结论 脑梗死患者血中CMV-IgG、CMV-IgM和CD-IgG升高,循环复合物对脑梗死的发生有一定的不良

    同类产品报价

    产品名称
    产品价格
    公司名称
    报价日期
    ¥1500
    上海抚生实业有限公司
    2025年07月12日询价
    ¥1200
    上海机纯实业有限公司
    2025年07月02日询价
    询价
    上海圻明生物科技有限公司
    2025年08月14日询价
    ¥1600
    瑟瑞娜生命科学技术发展(上海)有限公司
    2025年11月28日询价
    大鼠免疫循环复合物(rat CIq) ELISA试剂盒
    ¥2800