Innovative Bottle-Top Filters for Fast Vacuum Filtration of 100 mL to 5 L of Aqueous Solutions
- Filter faster. Supor® membrane provides high flow rates.
- No need to constantly refill upper fluid reservoir.
- Draws directly from the mixing reservoir.
- Eliminates the possibility of contamination from transfer steps. Filters directly into the desired container.
- Match filter to sample volume by choosing from two sizes.
- Available built-in prefilter increases throughput of particulate-laden solutions such as serum-containing media.
- Environmentally-friendly with minimal plastic waste.
- Patented small design accepts a variety of collection vessels and reduces storage space and waste.
- TA version available with tubing attached to each device for maximum convenience.
- Gamma irradiated, eliminates potential contamination by EtO residuals.
-
Materials of Construction
- Filter Media: Supor [hydrophilic polyethersulfone (PES)] membrane
- Housing: Modified acrylic
- Membrane Support Material: Polyester
- Sinker Material: Glass-filled polyurethane elastomer
- Inlet Tubing: Polyvinyl chloride (PVC) medical-grade tubing
- Gasket Seal Material: Polyethylene
-
Pore Size
- 0.1, 0.2, 0.45, and 0.8/0.2 μm
-
Effective Filtration Area
- VacuCap 60 Devices: 30 cm2
- VacuCap 90 Devices: 60 cm2
-
Typical Throughput
- (RPMI + 10% newborn calf serum)
VacuCap 60 Devices, 0.2 μm: 1 L
VacuCap 90 Devices, 0.2 μm: 5 L - (RPMI + 10% calf serum)
VacuCap 60 PF Devices: 500 mL
VacuCap 90 PF Devices: 1 L -
Collection Vessel
- VacuCap 60 Devices: Can be used on receptacles with openings ranging from 2 - 5 cm (0.8 - 1.9 in.)
- VacuCap 90 Devices: Can be used on receptacles with openings ranging from 2 - 6.5 cm (0.8 - 2.5 in.)
- Warning: Collection vessels should be vacuum rated. Implosion may result.
Typical Hold-Up Volume
- Vacucap 60 Devices: 1.2 mL
- VacuCap 90 Devices: 3.2 mL
-
Maximum Operating Temperature
- 55 °C (131 °F)
-
Typical Water Flow Rate
- mL/min at 25.4 cm Hg (10 in. Hg)
VacuCap 60 Devices VacuCap 90 Devices 0.1 µm: 50 0.1 µm: 100 0.2 µm: 200 0.2 µm: 400 0.45 µm: 280 0.45 µm: 560 PF: 200 PF: 400 -
Recommended Vacuum
- 38.1 cm Hg (15 in. Hg)
-
Endotoxin Level
- < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
-
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
-
Sterilization
- Sterilized by gamma irradiation and individually packaged
-
Part Number Description Pkg VacuCap 60 Devices 4631 0.1 µm, 60 mm, gamma-irradiated 10/pkg 4632 0.2 µm, 60 mm, gamma-irradiated 10/pkg 4634 0.45 µm, 60 mm, gamma-irradiated 10/pkg TA4632 0.2 µm, 60 mm (supplied with individually attached tubing for each filter device), gamma-irradiated 10/pkg VacuCap 60 PF Device 4638 0.8/0.2 µm, 60 mm, gamma-irradiated 10/pkg VacuCap 90 Devices 4621 0.1 µm, 90 mm, gamma-irradiated 10/pkg 4622 0.2 µm, 90 mm, gamma-irradiated 10/pkg 4624 0.45 µm, 90 mm, gamma-irradiated 10/pkg TA4622 0.2 µm, 90 mm (supplied with individually attached tubing for each filter device), gamma-irradiated 10/pkg VacuCap 90 PF Device 4628 0.8/0.2 µm, 90 mm, gamma-irradiated 10/pkg Accessories and Replacement Parts 4623 Feedline accessory kit 1/pkg
风险提示:丁香通仅作为第三方平台,为商家信息发布提供平台空间。用户咨询产品时请注意保护个人信息及财产安全,合理判断,谨慎选购商品,商家和用户对交易行为负责。对于医疗器械类产品,请先查证核实企业经营资质和医疗器械产品注册证情况。
文献和实验平均值应至少为0.5m/s。70%气体通过HEPA过滤器再循环至工作区,30%的气体通过排气口过滤排除。二级B型生物安全柜均为连接排气系统的安全柜。连接安全柜排气导管的风机连接紧急供应电源,目的在断电下仍可保持安全柜负压,以免危险气体泄漏如实验室。其前窗气流速度最小量或测量平均值应至少为0.5m/s(100fpm)。B1型70%气体通过排气口HEPA过滤器排除,30%的气体通过供气口HEPA过滤器再循环至工作区。B2型为100%全排型安全柜,无内部循环气流,可同时提供生物性和化学性的安全控制。三级
见参考文献,也可参考生产商的说明手册。 图8 Ⅱ 级B1型生物安全柜原理图 A:前开口;B:窗口;C:排风HEPA过滤器;D:供风HEPA过滤器; E:负压压力排风系统;F:风机;G:送风HEPA过滤器。 安全柜需要有与建筑物排风系统相连接的排风接口。 表9Ⅰ级、Ⅱ级以及Ⅲ 级生物安全柜之间的差异 (NA,不适用) a所有生物学污染的管道均为负压状态,或由负压的管道和压力通风系统围绕。Ⅲ级生物安全柜 Ⅲ级生物安全柜(图9)用于操作危险度4级的微生物材料,可以提供最好的个体
安全柜前窗气流速度最小量或测量平均值应至少为0.38m/s。70%气体通过HEPA过滤器再循环至工作区,30%的气体通过排气口过滤排除。A2型安全柜前窗气流速度最小量或测量平均值应至少为0.5m/s。70%气体通过HEPA过滤器再循环至工作区,30%的气体通过排气口过滤排除。A2型安全柜的负压环绕污染区域的设计,阻止了柜内物质的泄漏。 二级B型生物安全柜均为连接排气系统的安全柜。连接安全柜排气导管的风机连接紧急供应电源,目的在断电下仍可保持安全柜负压,以免危险气体泄漏如实验室。其前窗气流速度最小