        Optimization of Sensitivity

        互联网

        482
        The injection volumes involved in CE are very small (in the order of 1–10 nL). In addition, the area of capillary employed for on-column detection may be only 50 � 200 �m. Both these factors influence the detector sensitivity to a large extent. CE is less sensitive when directly compared to HPLC with typical injections of 10–50 �L and 1-cm detector cells. The difference may be up to an order of magnitude (2 ) when comparing at the same UV wavelength. Several strategies may be employed (Table 1 ) to maximize the sensitivity in CE. These include use of low UV wavelengths, increased capillary bore, and optimized sampling procedure.
        Table 1  Summary of Approaches Available for Increasing Sensitivity

        Action to improved sensivity

        Drawback(s) to consider

        Employ low-UV wavelength

        Increased background noise—determine wavelength to give optimum signal-to-noise ratio

        Increase capillary bore

        Increased current, reduced EOF giving possible alteration in selectivity

        Increased injection time

        Reduction in separation efficiency; excessive time will result in run failure

        Appropriate use of electrokinetic injection

        Sampling bias for more mobile ions, sample matrix effects on injection amount

        Increased electrolyte strength

        Increased current and associated noise

        Optimize electrolyte composition

        Effects on selectivity and current

        Decrease operating voltage

        Increase in analysis time

        Decrease temperature

        Increase in analysis time

        Capillary modifications

        Reduction in separation efficiency and resolution, increased cost

        Sample derivatization

        Additional sampling handling

        Indirect detection

        Extra method development considerations

        Wide-bore capillaries

        EOF profile disturbed, adjustments to rinse and injection times, siphoning effects more pronounced

        Increased detector slit width

        Reduction in separation efficiency and resolution