Parallel-plate Flow Chamber and Continuous Flow Circuit to Evaluate Endothelial Progenitor Cells under Laminar Flow Shea
互联网
实验材料
Cell-Culture Dishes (4-well, rectangular) |
|||
Pump-Drive (Masterflex L/S variable-speed economy drive) |
|||
Pump-Head (Masterflex Easy Load Pump Head) |
|||
Sytox Orange Nucleic Acid Stain |
|||
实验步骤
1. Endothelial progenitor cell isolation
1) Prior to any collection of peripheral human blood, submit your research protocol to your Institutional Review Board (IRB), and after its approval, obtain the volunteer donors' informed consent (peripheral blood collection and EPC isolation had been approved by the Duke University IRB and is in full compliance with U.S. regulatory requirements related to the protection of human research participants).
2) When working with animal-derived EPCs, have your research protocol approved by your Institutional Animal Care and Use Committee (IACUC). All our porcine experiments had been approved by the Duke University IACUC and were conducted in accordance with the highest standards of humane care.
3) For isolation of endothelial progenitor cells, collect 50 ml of peripheral blood via standard phlebotomy technique from a consented volunteer donor into blood collection bags filled with the anticoagulant citrate phosphate dextrose and dilute the solution 1:1 with Hank's buffered salt solution (without CaCl2 , MgCl2 , MgSO4 ) and layer on equal volumes of Histopaque to create well-defined layers.
4) Centrifuge (30 min, 740 g, low break setting) and collect the mononuclear cell (MNC) (buffy coat) layer. Resuspend and wash MNCs x 3 with Dulbecco's Phosphate Buffered Saline (DPBS) with 10% Fetal Bovine Serum (FBS) x 10 min, 515 g, before plating into two 12-well plates in full EPC growth medium (MCDB-131 medium with 5 ml of 200 mM L-glutamine, 10% FBS and EGM-2 SingleQuots) at 37 °C, 5% CO2 .
5) Slowly change medium every 24 hours for the first 7 days, then every other day. EPC colonies can be identified after an average time of 14 days based on their cobblestone morphology. Once EPCs in culture cover ¼ of the 12-well surface area, expand cells and confirm EPC identity with flow cytometry by testing for presence of surface markers CD31 and absence of CD14, CD45 as previously described 6 . Other assays that may be performed include cell morphology and labeling with DiI-Ac-LDL.
6) For the experiments described herein, expand isolated EPCs until they are nearly confluent in 3 T-75 cell culture flasks in EPC medium in a humidified incubator at 37 °C, 5% CO2 .
1) We recommend that these calculations are performed before manufacturing the chamber and channel to have a thorough understanding of the conditions that can be achieved. Calculate the flow rate of the perfusate in your circuit based on the desired shear stresses to be applied to the cells according to equation 1 1 , where Q is the desired flow rate, τ is the target shear stress acting tangentially on the cells, w is the width of the flow chamber, h is the height of the flow chamber, and μ is the viscosity of the perfusate (flow medium).
2) A typical value of viscosity (μ ) for medium used is 0.9 cP (0.009 g cm-1 s-1 ). Note that the viscosity can also be raised by using high molecular weight dextran, if so desired 7 . A typical value of target shear stress (τ ) used is 15 dynes/ cm2 , (typical arterial shear stress) or 100 dynes/cm2 , if supra-physiological shear stress is desired. Our chamber typically has a width (w ) of 1.9 cm and height (h ) in the range of 166 - 267 μm.
1) Cut the top and bottom plates of the chamber from aluminum alloy 6061 rectangular stock with top dimension of 7 x 2 x 0.5" and bottom dimensions of 7 x 2 x 0.25".
2) Recess the top plate in the center portion to 0.125" deep, leaving 0.950" per end for 1.375" wide x 0.3125" deep fluid reservoir.
3) Drill 1/16" wide x 0.625" long slots from the underside in order to penetrate to the reservoir on the inflow and outflow ends.
4) At the end of each top plate, drill a centrally located 10/32" tracks per inch (T.P.I.) hole for polypropylene 1/8" hose barb to penetrate the fluid reservoir. On the inflow end, create a side plug with 10/32" T.P.I through hole for polypropylene 1/8" hose barb.
5) Taper the top underside from the fluid inflow to the glass viewing window at 0.518 degrees for adequate mixing of fluid.
6) Cut a glass slide window into the bottom plate in line with the top viewing window. Using aquarium cement, glue a glass slide into the chamber. Allow to cure for at least 24 hours. Ensure that the viewing window is recessed so that a 75 x 25 x 1 mm microscope glass slide will sit flush in the chamber.
7) Create a recess for an O-ring around the underside of the top and bottom plates such that they are spaced apart by 0.02". Insert the rubber O-rings.
8) Drill 10 matching 8/32" T.P.I. holes in top and bottom chambers for use with 8/32" flat head screws.
1) Assemble the entire flow circuit first, then proceed with its sterilization.
2) Attach a 36" segment of hard tubing to one end of the pulse dampener (via 1/8" connection). To the other end, attach an 18" segment of soft tubing.
3) Place a 1/8" male luer adaptor at the end of the 18" soft tubing section, mentioned in step 4.2. Connect the male luer to a 1/8" female luer adaptor. Attach the female luer to a new 18" segment soft tubing.
4) Drill three holes into a 250 ml glass beaker cap in order to fit tubing. Insert a 1.5" segment of soft tubing (as air vent) into one hole and the free ends of the hard and soft tubing through the other 2 holes in the cap into the 250 ml glass bottle (as reservoir) (Fig. 2). Ensure that the hard tubing reaches the bottom of the reservoir (as outflow tubing from the reservoir).
5) Sterilize the above parts by steam autoclaving at 121 °C for 60 min. Additionally autoclave one complete flow chamber, 3 x 2" segments of soft tubing, 1 male and 1 female 1/8" luer adaptor, 4 x ½" and 6 x 5/16" 8-32 flat head screws, 1 pair of tweezers, one 75 x 25 x 1 mm glass slide (or other desired cell surface material), and 2 surgical towels.
6) Place sterile towels into laminar flow hood. Place the flow circuit, flow chamber, 4 x 4-way stopcocks, 1 x 1-way stopcock, and all the sterilized parts from step 4.5 onto this sterile field.
7) Now use sterile gloves to connect two 4-way stopcocks. Place caps on the open sample ports.
8) Detach the male and female luer adaptors attaching the two soft tubing segments of the flow circuit and insert the connected stopcocks (Fig. 3).
9) Attach the soft tubing inserted into the reservoir to a 30 ml syringe and remove the syringe piston. Fill the bottle with 125 ml of EPC medium (Fig. 4).
10) Place a sterile syringe filter into the air vent (Fig. 4).
11) Move the flow circuit into a humidified incubator at 37 °C, 5% CO2 .
12) Clamp the hard tubing into the roller pump head indicated for use with this type of tubing (Fig. 5). Ensure that the tubing does not overlap with the roller pump mounting tracks. Mark the tubing where it exits the pump head on either side with a marking pen.
13) Ensure that all stopcocks are closed off towards the sample ports and open along the flow circuit. Start the pump. Verify the direction of flow. The perfusate should move from the glass reservoir through the hard tubing into the pulse dampener.
1) While the flow circuit warms to 37 °C, prepare the cells for seeding onto the slide. Note that fluorescent labeling is required if cells are to be visualized on opaque materials, e.g. titanium (Ti).
2) Create a 1 mM stock solution of Cell Tracker Orange (CTO) by diluting 50 μg CTO powder with 90 μl of dimethyl sulfoxide (DMSO). Be sure to shield this mixture from light exposure (turn off light while working under flow hood).
3) Create a 2 μM working solution of CTO by diluting 36 μl of CTO stock solution into 18 ml of serum-free MCDB-131 medium.
4) Rinse EPCs in 3 near confluent T-75 cell culture flasks with 10 ml DPBS (without Ca or Mg).
5) Add 6 ml of the 2 μM CTO working solution to each flask. Incubate for 15 min at 37 °C. Rinse each flask x 2 with 10 ml DPBS (without Ca or Mg).
6) Add 4 ml of Trypsin to each flask. Incubate at 37 °C for 3 min. Neutralize with 8 ml of Trypsin Neutralization Solution.
7) Wash by centrifugation x 5 min at 600g and resuspend in 5 ml of EPC medium.
6. Cell seeding of slide prior to flow chamber assembly
1) A slide made out of glass, titanium, or other material can be used, and the slide surface can be modified by spin coating with polymer or adding a layer of metal. Place the slide into a sterile rectangular four-chamber cell culture vessel (or use tissue culture slide flask if polystyrene surface is desired). Count cells and seed 1.5 x 106 cells onto the slide in 5 ml of EPC medium x 6 hr if a confluent cell layer is desired at the initiation of flow (Fig. 11A).
2) In order to test cell spreading and adhesion under fluid shear stress, allow 3 x 106 cells to adhere for only 15 min before commencement of flow (quick-seeding) (Fig. 11B).
1) Using sterile gloves, connect one 2" soft tubing segment to a 1/8" female luer adaptor. Connect another 2" soft tubing segment to a 1/8" male luer adaptor.
2) Attach the 2" soft tubing with female luer adaptor to the inflow port. Attach the 2" soft tubing with male luer to the outflow port. Attach 4-way stopcocks to both of these luer adaptors. Connect the remaining 2" soft tubing piece to the side port connector coming off the bubble trap and attach a 1-way stopcock (Fig. 6).
3) Using sterile tweezers, remove the cell-seeded slide from the cell culture vessel (or if using a slide flask, remove the flask on slide) and place it into the recess on the bottom plate of the flow chamber. Make sure that the slide is placed with the cell-seeded side facing up.
4) Pipet 10 ml of warm EPC medium onto the slide. Allow the medium to cover the slide and flow chamber, but do not spill the medium over the O-ring on the bottom plate.
5) Place the top plate of the flow chamber onto the bottom plate, aligning the screw holes. Take care not to introduce air bubbles into the chamber during this step by keeping the two plates parallel.
6) Screw plates together (an automatic battery-operated screwdriver speeds up this process).
7) Remove air bubbles from inflow bubble trap by opening the 1-way stopcock attached to the inflow-side bubble trap port and gently flush the inflow tubing with EPC medium using a 10 ml syringe. Then close off this stopcock and cap it (Fig. 7).
8) Now remove air bubbles from the chamber channel by opening the outflow port 4-way stopcock and by gently flushing EPC medium through the flow chamber, again using a 10 ml syringe. If large bubbles remain, raise the outflow end of the chamber to a 45° angle (Fig. 8).
9) Cap all of the 4-way stopcock connections and close them off. Move the flow chamber into the incubator with the assembled flow circuit ( Fig. 9).
10) Pause the flow circuit pump. Close off the flow circuit stopcocks in the direction of flow to prevent leakage and to keep the inside of the chamber sterile. Transport the flow chamber from the flow hood to the flow circuit. Connect the flow chamber to the flow circuit. Open the stopcocks in the direction of flow. Resume the flow pump. Ensure that the perfusate flows in the correct direction. Adjust the flow rate to the desired shear stress.
11) During the flow experiment, the flow chamber can be removed from the circuit to image the cells via light or fluorescent microscopy. In order to do so, disconnect the flow chamber from the flow circuit at the stopcock-stopcock connections. To maintain sterility of the chamber, close off the flow circuit stopcocks and flow chamber stopcocks in the direction of flow and cap all stopcocks prior to removing the flow circuit from the incubator (Fig. 9).
8. Perfusate sample collection
1) For easy collection of perfusate samples for analysis, (e.g. nitric oxide synthesis), first pause the pump.
2) Close the stopcock located closest to the pulse dampener. Remove its protective cap and insert a small size syringe, e.g. 3 ml syringe, into the sample port. Keep cap and ensure that it will not become contaminated by placing it upside down.
3) Close the stopcock in the direction of the flow chamber, keeping the sample port and circuit in the direction of the pulse dampener open.
4) Draw desired amount of sample, e.g. 150 μl, from the circuit. Close off the stopcock towards the sample port before removing the syringe. Store the perfusate sample in a labeled vial and freeze at - 80 °C (for nitric oxide determination at a later time point).
5) Recap the sample port and ensure that all stopcocks are open before commencing flow.
Using our quick-seed method, human blood-derived endothelial progenitor cells can be seeded onto titanium slides for 15 minutes and adhere under physiological (arterial) shear forces of 15 dynes/cm2 .
As shown in Figure 10, EPCs spread under the influence of flow from 210 ± 11.4 μm2 , immediately after seeding, to 657 ± 39.1 μm2 after 3 hours, and to 1152 ± 55.3 μm2 after 24 hours of fluid shear stress of 15 dynes/ cm2 (difference between the groups is statistically significant with p < 0.0001 , 1-way ANOVA). Parallel to the increase in cell area, which can be imaged through the transparent chamber with either light or fluorescent microscope, the roundness calculated according to equation 2 6
decreased from 878 x 10-3 ± 5.9 x 10-3 , immediately after seeding, to 671 x 10-3 ± 19.2 x 10-3 after 3 hours, and to 526 x 10-3 ± 19.2 x 10-3 after 48 hours of the same fluid shear stress exposure (difference between the groups is again statistically significant with p < 0.0001 , 1-way ANOVA).
Figure 11D illustrates that EPCs align and orient in the direction of flow after 48 hours of fluid shear stress at 15 dynes/cm2 as compared to their random orientation after a static seeding period of 6 hours, shown in Figure 11A.
Our method also allows for obtaining samples of the perfusate at predetermined time points for the analysis and/or quantification of secreted cell metabolites. As a representative example we depict the production of nitrite (NO2 -) during a 48 hour flow experiment with porcine EPCs in Figure 12. We directly measured this primary oxidation product of nitric oxide (NO) as a surrogate marker for NO in medium samples collected from the flow circuit at the different time points to be 12.0 nmol at 6 hours, 13.1 nmol at 12 hours, 16.3 nmol at 24 hours, and 24.6 nmol after 48 hours for 1 x 106 cells at 15 dynes/cm2 6 .
1. Bhat, V.D., Truskey, G.A., & Reichert, W.M. Fibronectin and avidin-biotin as a heterogeneous ligand system for enhanced endothelial cell adhesion. J. Biomed. Mater. Res. 41, 377-385, doi:10.1002/(SICI)1097-4636(19980905)41:3<377::AID-JBM6>3.0.CO;2-9 [pii] (1998).
2. Broxmeyer, H.E., et al. Cord blood stem and progenitor cells. Methods Enzymol. 419, 439-473, doi:S0076-6879(06)19018-7 [pii] 10.1016/S0076-6879(06)19018-7 (2006).
3. Achneck, H.E., et al. In American Heart Association Scientific Sessions, Abstract Oral Sessions, Medical Aspects End Stage Heart Failure: Transplantation and Device Therapies.
4. Brown, M.A., Wallace, C.S., Angelos, M., & Truskey, G.A. Characterization of umbilical cord blood-derived late outgrowth endothelial progenitor cells exposed to laminar shear stress. Tissue. Eng. Part. A. 15, 3575-3587, doi:10.1089/ten.TEA.2008.0444 (2009).
5. Achneck, H.E., et al. Regenerating titanium ventricular assist device surfaces after gold/palladium coating for scanning electron microscopy. Microsc. Res. Tech. 73, 71-76, doi:10.1002/jemt.20757 (2010).
6. Achneck, H.E., et al. The biocompatibility of titanium cardiovascular devices seeded with autologous blood-derived endothelial progenitor cells: EPC-seeded antithrombotic Ti Implants. Biomaterials. 32, 10-18, doi:S0142-9612(10)01111-7 [pii] 10.1016/j.biomaterials.2010.08.073 (2011).
7. Rinker, K.D., Prabhakar, V., & Truskey, G.A. Effect of contact time and force on monocyte adhesion to vascular endothelium. Biophys. J. 80, 1722-1732, doi:10.1016/S0006-3495(01)76143-7 (2001).
8. Truskey, G.A., Yuan, F., & Katz, D.F. In Transport Phenomena in Biological Systems. , Ch. 2, Horton, M.J., ed., Pearson Education, (2004).
9. Truskey, G.A., Yuan, F., & Katz, D.F. In Transport Phenomena in Biological Systems. , Ch. 2, Horton, M.J., ed., Pearson Education, (2009).
10. Allen, J.D., et al. Plasma nitrite response and arterial reactivity differentiate vascular health and performance. Nitric Oxide. 20, 231-237, doi:S1089-8603(09)00011-1 [pii] 10.1016/j.niox.2009.01.002 (2009).
11. Xiao, Y. & Truskey, G.A. Effect of receptor-ligand affinity on the strength of endothelial cell adhesion. Biophys. J. 71, 2869-2884, doi:10.1016/S0006-3495(96)79484-5 (1996).
12. Conant, C.G., Schwartz, M.A., Nevill, T., & Ionescu-Zanetti, C. Platelet Adhesion and Aggregation Under Flow using Microfluidic Flow Cells. J. Vis. Exp. (32), e1644, DOI: 10.3791/1644 (2009).
13. Conant, C.G., Schwartz, M.A., & Ionescu-Zanetti, C. Well plate-coupled microfluidic devices designed for facile image-based cell adhesion and transmigration assays. J. Biomol. Screen. 15, 102-106, doi:10.1177/1087057109353789 (2010).
14. Conant, C.G., et al. Using well-plate microfluidic devices to conduct shear-based thrombosis assays. J. Lab. Autom. 16, 148-152, doi:10.1016/j.jala.2010.10.005 (2011).
15. Metaxa, E., et al. Nitric oxide-dependent stimulation of endothelial cell proliferation by sustained high flow. Am. J. Physiol. Heart. Circ. Physiol. 295, H736-742, doi:10.1152/ajpheart.01156.2007 (2008).
16. Radel, C., Carlile-Klusacek, M., & Rizzo, V. Participation of caveolae in beta1 integrin-mediated mechanotransduction. Biochem. Biophys. Res. Commun. 358, 626-631, doi:10.1016/j.bbrc.2007.04.179 (2007).
17. Wang, X.L., Fu, A., Spiro, C., & Lee, H.C. Proteomic Analysis of Vascular Endothelial Cells-Effects of Laminar Shear Stress and High Glucose. J. Proteomics. Bioinform. 2, 445, doi:10.4172/jpb.1000104 (2009).
18. Elfervig, M.K., Minchew, J.T., Francke, E., Tsuzaki, M., & Banes, A.J. IL-1beta sensitizes intervertebral disc annulus cells to fluid-induced shear stress. J. Cell. Biochem. 82, 290-298 (2001).
19. Archambault, J.M., Elfervig-Wall, M.K., Tsuzaki, M., Herzog, W., & Banes, A.J. Rabbit tendon cells produce MMP-3 in response to fluid flow without significant calcium transients. J. Biomech. 35, 303-309 (2002).
20. Patton, J.T., Menter, D.G., Benson, D.M., Nicolson, G.L., & McIntire, L.V. Computerized analysis of tumor cells flowing in a parallel plate chamber to determine their adhesion stabilization lag time. Cell. Motil. Cytoskeleton. 26, 88-98, doi:10.1002/cm.970260109 (1993).
21. Wiese, G., Barthel, S.R., & Dimitroff, C.J. Analysis of Physiologic E-Selectin-Mediated Leukocyte Rolling on Microvascular Endothelium. J. Vis. Exp. (24), e1009, DOI: 10.3791/1009 (2009).
22. Kaper, H.J., Busscher, H.J., & Norde, W. Characterization of poly(ethylene oxide) brushes on glass surfaces and adhesion of Staphylococcus epidermidis. J. Biomater. Sci. Polym. Ed. 14, 313-324 (2003).
23. Bakker, D.P., van der Plaats, A., Verkerke, G.J., Busscher, H.J., & van der Mei, H.C. Comparison of velocity profiles for different flow chamber designs used in studies of microbial adhesion to surfaces. Appl. Environ. Microbiol. 69, 6280-6287 (2003).
24. Angelos, M.G., et al. Dynamic adhesion of umbilical cord blood endothelial progenitor cells under laminar shear stress. Biophys. J. 99, 3545-3554, doi:S0006-3495(10)01252-X [pii] 10.1016/j.bpj.2010.10.004 (2010).
25. Baker, G.R., Sullam, P.M., & Levin, J. A simple, fluorescent method to internally label platelets suitable for physiological measurements. Am. J. Hematol. 56, 17-25, doi:10.1002/(SICI)1096-8652(199709)56:1<17::AID-AJH4>3.0.CO;2-5 [pii] (1997).
注意事项
1. It is desirable to avoid bubble formation during the flow set-up, since bubbles have the potential to tear off cells from their surface. This can be avoided by taking care when placing the top part of the flow chamber onto the bottom part, which is best accomplished by keeping both parallel to each other and lowering the top onto the bottom in one motion without readjustments. In doing so, small air bubbles that may be present and/ or floating in the flow channel, will be diverted into the bubble traps on either end of the aluminum housing.
2. It is important to ensure that the outflow tubing in the reservoir reaches down to the reservoir's bottom. Otherwise, air may be sucked into the tubing that leads from reservoir to chamber if the perfusate level falls slightly below the tubing end.
3. The researcher should be familiar with the direction of the pump flow so that the pump does not accidentally run in the opposite direction upon commencement of flow, which would result in air being sucked into the circuit and chamber.
4. To avoid the formation of foam (due to denatured proteins contained in serum), we advise lowering the tubing that leads from chamber to reservoir to approximately one inch above the perfusate level inside the reservoir. However, keeping it above the flow medium level allows you to verify flow into the reservoir during the experiment.
5. When screwing the chamber parts together, it is important to firmly grasp the housing with one hand while operating the electric screwdriver with your other hand. Note that the motion of the electric screwdriver will be translated into torque which has the potential to 'fling the chamber around' once the screw is tightened.
6. To avoid any pressure waves from forming in the circuit and lifting cells off their surface, ensure that all stopcocks are open before commencing flow.
7. Especially for longer term flow studies (> 48 hours) we recommend using the harder and more resilient tubing to be inserted into the pump head to prevent breaking of tubing.
8. It is possible that tubing 'migrates' inside the pump head due to ill-adjusted pump head teeth. Therefore we suggest paying close attention to how the tubing is secured in the pump head and marking each end of the tubing with a marking pen such that you can easily notice if the tubing is 'pulled in' or dislodged during the experiment.
9. Always carefully check every single connector of the circuit and chamber prior to starting the perfusion in order to prevent leakage of perfusate during an experiment due to a faulty connection. This is especially important for the inflow and outflow connectors of pulse dampeners!
10. Remember to limit light exposure if you use fluorescent labels.