6PLUS1® Multilevel Whole Blood Calibrator Set Immunosuppressants
保存条件:
低温
规格:
7 x 2 ml
6PLUS1® Multilevel Whole Blood Calibrator Set Immunosuppressants
Order No.: 28039
参数:
环孢素A、依维莫司、西罗莫司、他克莫司
Description
6PLUS1® Multilevel Calibrator Whole Blood Set Immunosuppressants, lyophilised For MassTox® Immunosuppressants in Whole Blood (93000, 93000/1200)
These lyophilised calibrators from Chromsystems are based on human whole blood and should be used according to the instruction manual. They are designed according to the recommendations of the FDA to calibrate the measurement method in clinical diagnostics for the quantitative determination of cyclosporin A, tacrolimus, sirolimus and everolimus in whole blood. While following the instructions the calibrators should be handled and measured in the same manner as a patient specimen. These calibrators are only available as multilevel calibrator set with 6 different concentration levels plus 1 blank value.
Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.
Substance
Unit
Calibrator 1 (28031)
Calibrator 2 (28032)
Calibrator 3 (28033)
Calibrator 4 (28034)
Calibrator 5 (28035)
Calibrator 6 (28036)
Blank Calibrator (28030)
Cyclosporin A (Ciclosporin)
µg/l
25.3
125
301
476
800
971
n.d. (< 10)
Tacrolimus (FK 506)
µg/l
1.5
5.82
11.9
17.2
24.3
41.6
n.d. (< 0.5)
Rapamycin (Sirolimus)
µg/l
2.27
6.46
13.3
19.4
30.4
49.6
n.d. (< 1)
Everolimus
µg/l
2.31
6.03
12.2
17.7
25.4
44.1
n.d. (< 1)
n.d. = not detected (< LOQ)
The blank level is manufactured using the same matrix pool as used for the calibrator levels 1-6. It was not fortified with any immunosuppressants.
Storage conditions
Stored below -18 °C and unopened, the lyophilised calibrators are stable until the expiry date specified on the vial labels. The reconstituted calibrators can be stored up to 1 week light protected and tightly capped at +2 to +8 C. If the calibrators cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 9 months). For more details about storage conditions please check your instruction manual.