特异性: Natural and recombinant Human Protachykinin-1
精密度
Intra-assay Precision (Precision within an assay):Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Intra-Assay CV: ≤4.8%
Inter-assay Precision (Precision between assays):Three samples of known concentration were tested in five separate assays to assess inter-assay precision.
Inter-Assay CV: ≤8.9%
回收率
回收率:低、中和高浓度的分析物被掺入到血清或者血浆样本中,进行回收实验测定。
Sample Type
Average(%)
Recovery Range(%)
Serum
98
92-104
Plasma
100
94-106
线性
给定样本通过梯度稀释,每次稀释的测量值与理论值的比值。
Sample
1:2
1:4
1:8
1:16
serum(n=5)
80-90%
91-101%
102-114%
98-108%
EDTA plasma(n=5)
108-118%
117-127%
105-117%
106-116%
heparin plasma(n=5)
106-114%
101-111%
88-98%
97-105%