特异性: Natural and recombinant Human Angiotensin-converting enzyme
研究领域:Neurosciences
精密度
Intra-assay Precision (Precision within an assay):Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Intra-Assay CV: ≤5.7%
Inter-assay Precision (Precision between assays):Three samples of known concentration were tested in five separate assays to assess inter-assay precision.
Inter-Assay CV: ≤8.9%
回收率
回收率:低、中和高浓度的分析物被掺入到血清或者血浆样本中,进行回收实验测定。
Sample Type
Average(%)
Recovery Range(%)
Serum
99
93-105
Plasma
101
95-107
线性
给定样本通过梯度稀释,每次稀释的测量值与理论值的比值。
Sample
1:2
1:4
1:8
1:16
serum(n=5)
100-113%
107-117%
101-111%
102-115%
EDTA plasma(n=5)
113-122%
102-112%
96-106%
96-107%
heparin plasma(n=5)
104-114%
109-119%
87-97%
96-105%