Glo1 ELISA Kit allows for the in vitro quantitative determination of GLO1 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Matrices listed below were spiked with certain level of GLO1 and the recovery rates were calculated by comparing the measured value to the expected amount of GLO1 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-98
92
EDTA plasma(n=5)
89-95
92
heparin plasma(n=5)
85-98
90
线性度
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GLO1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.