IDUA ELISA Kit allows for the in vitro quantitative determination of IDUA concentrations in serum, plasma, tissue homogenates and other biological fluids.
Matrices listed below were spiked with certain level of IDUA and the recovery rates were calculated by comparing the measured value to the expected amount of IDUA in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-103
94
EDTA plasma(n=5)
90-104
96
heparin plasma(n=5)
87-104
92
线性度
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IDUA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.