βcellulin/BTC ELISA Kit allows for the in vitro quantitative determination of β-TC concentrations in serum, plasma, tissue homogenates and other biological fluids.
Matrices listed below were spiked with certain level of β-TC and the recovery rates were calculated by comparing the measured value to the expected amount of β-TC in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
90-105
98
EDTA plasma(n=5)
93-100
97
heparin plasma(n=5)
95-103
98
线性度
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of β-TC and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.