CXCR1 ELISA Kit allows for the in vitro quantitative determination of CXCR1 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Matrices listed below were spiked with certain level of CXCR1 and the recovery rates were calculated by comparing the measured value to the expected amount of CXCR1 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
87-104
95
EDTA plasma(n=5)
91-104
99
heparin plasma(n=5)
89-105
99
线性度
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CXCR1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.